High Temperature Cables TEBAOFLEX

Pt100 & Pt1000 Screened FEP & PFA RTD Sensor Cable (3-, 4- & 6-Wire)

Pt100 and Pt1000 screened FEP/C/FEP (-90 °C to +205 °C) and PFA/C/PFA (-90 °C to +260 °C) RTD sensor connection cables (TF-FEP-PFA-PT100-RTD) feature resistance-matched plated copper conductors (0.14 mm² to 1.50 mm²), IEC 60751 red/white polarity cores and >= 85 % plated copper EMC braid screening. Built for 3-wire, 4-wire Class AA/A and duplex 6-wire platinum resistance thermometers on 300 m reels.

Product information
SKU
TF-FEP-PFA-PT100-RTD
Brand
TEBAOFLEX
Product overview

Screened FEP/C/FEP and PFA/C/PFA Pt100, Pt500 and Pt1000 resistance thermometer connection cables deliver ultra-stable milliohm signal transmission from -90 °C cryogenic skids up to +260 °C process thermowells, spanning 2.8 mm miniature 4×0.14 mm² probe transition tails up to 7.4 mm 4×1.50 mm² field transmitter leads.

On a 100 Ohm Pt100 sensor (0.385 Ohm/°C temperature coefficient), just 0.19 Ohm of uncompensated lead resistance imbalance or wet-insulation leakage creates a 0.5 °C error — enough to push an IEC 60751 Class A (+/-0.15 °C) or Class AA (+/-0.10 °C) sensor out of specification. TEBAOFLEX pairs co-drawn tinned, silver-plated or nickel-plated copper strands inside > 5 GOhm*km fluoropolymer insulation and a >= 85 % copper EMC braid. Low-capacitance fluoropolymer dielectrics and concentric star-quad core geometry also suppress triboelectric and electromagnetic transient noise near variable-speed drives.

Select 3-wire builds (2 red + 1 white) for general industrial transmitters, 4-wire Kelvin builds (2 red + 2 white) for Class AA custody-transfer, pharmaceutical SIP and semiconductor chillers, or 6-wire builds (4 red + 2 white) for redundant dual-element Pt100 probes (MOQ 300 m per construction, 7 to 45 working days).

Explore the high-temperature wire and cable range for related configurations.

Construction

Pt100 FEP/C/FEP & PFA/C/PFA RTD Cable Construction

All cores in a 3-, 4- or 6-wire RTD lot are stranded from the same wire-drawing batch and laid symmetrically to equalise DC resistance and thermal EMF.

Clean 2D engineering cross-section of 4-wire screened FEP/C/FEP Pt100 RTD cable showing outer sheath, silver-plated copper EMC braid, PTFE separator tape and 2 red plus 2 white cores
4-wire star-quad RTD cross-section: symmetrical resistance-matched plated copper strands, thin-wall FEP/PFA insulation, PTFE wrap and EMC braid.
Pt100 & Pt1000 Screened FEP & PFA RTD Sensor Cable (3-, 4- & 6-Wire) construction
ComponentMaterial and construction
1. Matched Plated Copper Strands (0.14 mm² to 1.50 mm²)Fine 7-strand or 19-strand tinned copper (up to +180 °C), silver-plated copper (up to +205/+250 °C) or nickel-plated copper (up to +260 °C) with < 1.5 % inter-core resistance spread.
2. Thin-Wall FEP or PFA Core Insulation (IEC 60751 Colours)Pinhole-free fluoropolymer core wall (> 5 GOhm*km) colour-coded per IEC 60751: red + white (2-wire), 2 red + 1 white (3-wire), 2 red + 2 white (4-wire) or 4 red + 2 white (6-wire).
3. Symmetrical Lay-Up, PTFE Wrap & >= 85 % Plated Copper BraidSymmetric star-quad (4-wire) or concentric lay-up wrapped with low-friction PTFE tape and an >= 85 % optical-coverage plated copper EMC screen.
4. Smooth Extruded FEP or PFA Outer SheathRust-brown, white or black cleanroom-grade fluoropolymer outer sheath impervious to CIP/SIP caustic washdowns, steam, solvents, acids and oils.

Pt100 IEC 60751 RTD Cable Code & Colour Breakdown

  • FEP / PFA: Core and outer sheath polymer: FEP (-90 to +205 °C) or PFA (-90 to +260 °C)
  • C: Overall >= 85 % tinned, silver-plated or nickel-plated copper wire EMC braid
  • 3x / 4x / 6x: 3-wire (2R+1W), 4-wire Kelvin (2R+2W) or 6-wire duplex Pt100 (4R+2W) per IEC 60751
  • 0.14–1.50 mm²: Matched plated-copper conductor cross-section (26 AWG to 16 AWG)
  • SPC / NPC: Silver-plated copper (<= +205/+250 °C) or nickel-plated copper (<= +260 °C)
Applications

Pt100 CIP/SIP Bioreactor, Semiconductor & Cryogenic Duty

Purpose-engineered for direct Pt100 / Pt1000 probe potting transitions and field thermowell-to-transmitter wiring.

Rust-red fluoropolymer screened Pt100 RTD sensor cables connected via stainless compression glands to sanitary thermowell heads on clean-steam process piping
Autoclave- and CIP/SIP-resistant fluoropolymer RTD connection cables wired to spring-loaded Pt100 thermowells.
  • Sanitary pharmaceutical bioreactors, WFI clean-steam loops, CIP/SIP skids and food pasteurizers requiring waterproof, autoclave-proof 4-wire Pt100 leads.
  • Semiconductor wafer-chuck chillers (-90 °C to +200 °C), wet chemical etch baths and ultra-clean PWIS/silicone-free tool harnesses.
  • Custody-transfer LNG/petrochemical metering skids, turbine bearing temperature probes and Stator Winding RTD slot leads.
  • Laboratory calibration baths, climatic chambers, vacuum ovens and industrial plastics extruder barrel sensors.

FEP/C/FEP & PFA/C/PFA Pt100 Cable vs Silicone SiHFC

FEP/C/FEP & PFA/C/PFA RTD Cable vs Silicone SiHFC vs Thermocouple Extension Cable
Cable ConstructionConductor MetalTemp. & OD AdvantageTarget Sensor Type
FEP/C/FEP RTD (this page)Silver/tinned copper-90 to +205 °C; 3.1 mm slim OD on 4x0.22 mm²Pt100 / Pt1000 Class AA/A probes, CIP/SIP & cleanrooms
PFA/C/PFA RTD (this page)Silver/nickel-plated copper-90 to +260 °C continuous chemical/vacuum duty260 °C stator slot RTDs, semiconductor & acid reactors
Silicone SiHFC / FEP-BiTinned copper-60 to +180 °C; thicker soft rubber wallGeneral dry ovens where PWIS/solvent immunity is not required
Thermocouple KX/TX (FLU06)NiCr-NiAl / Cu-CuNi alloysHigh loop resistance (never use on Pt100!)Millivolt thermocouple sensors only (IEC 60584)

For cable runs > 15 m or Class AA/A accuracy, specify 4-wire RTD construction (0.34 mm² to 1.0 mm²) to eliminate lead-wire resistance error completely.

Pt100 & Pt1000 Fluoropolymer RTD Cable RFQ Checklist

  • RTD wiring configuration: 3-wire (2 red + 1 white), 4-wire Kelvin (2 red + 2 white), 6-wire duplex (4 red + 2 white) or 8-wire dual 4-wire.
  • Conductor cross-section (0.14, 0.22, 0.34, 0.50, 0.75, 1.00 or 1.50 mm²) matched to probe sleeve ID or transmitter terminal gauge.
  • Temperature grade and plating: FEP/C/FEP (+180 °C TPC or +205 °C SPC) or PFA/C/PFA (+250 °C SPC or +260 °C NPC).
  • Outer sheath colour (rust-brown, white cleanroom or black) and reel quantity (MOQ 300 m per size, 7 to 45 working days).

Related selection resources

Pt100 Screened FEP & PFA RTD Sensor Cable FAQ

Should I choose 3-wire or 4-wire screened FEP/PFA cable for Pt100 sensors?

4-wire (2 red + 2 white) is mandatory for IEC 60751 Class AA (+/-0.10 °C) and Class A (+/-0.15 °C) loops because separate current-excitation and voltage-sense leads cancel lead resistance 100 %, whereas 3-wire assumes all three conductors have identical resistance.

Why is silver-plated copper preferred over nickel-plated copper below +205 °C on Pt100 leads?

Lower DC resistance and clean solderability: silver-plated copper has ~4 % lower resistance than nickel-plated copper and wets effortlessly with lead-free solder inside miniature 3 mm–6 mm stainless Pt100 probe transitions; reserve nickel-plated copper for +250 °C to +260 °C PFA crimp/weld terminations.

Why does insulation resistance (> 5 GOhm*km) matter so much on Pt1000 sensors?

Parallel shunt error: at +250 °C a Pt1000 element measures ~1,941 Ohm, so wet silicone or fibreglass insulation dropping to 1 MOhm shunts the element and causes an immediate -0.5 °C reading error; non-hygroscopic FEP and PFA stay orders of magnitude higher.

Are these FEP/C/FEP and PFA/C/PFA RTD cables silicone-free for semiconductor and paint plants?

Yes — 100 % fluoropolymer core and sheath extrusion uses zero silicone oils or mould-release agents (PWIS/LABS-free), making them safe for automotive e-coat ovens and cleanrooms.

What is the minimum order quantity and lead time for Pt100 FEP/C/FEP & PFA cables?

300 m per cross-section and temperature grade in continuous factory reels, manufactured and routine dielectric/resistance tested in 7 to 45 working days.

Specifications

FEP/C/FEP (205 °C) & Pt100 PFA/C/PFA (260 °C) Schedules

Pt100 Loop Resistance, Insulation Resistance & Reel Summary

  • Resistance-matched RTD cores: Tightly balanced plated copper stranding (0.14 mm² to 1.50 mm²) minimizes lead-wire bridge unbalance on 3-wire and 4-wire Class AA / Class A Pt100 loops
  • IEC 60751 red & white coding: Standardised red/white core combinations (2 red + 1 white for 3-wire; 2 red + 2 white for 4-wire; 4 red + 2 white for duplex 3-wire RTDs)
  • FEP 205 °C & PFA 260 °C grades: Cryogenic -90 °C up to +205 °C (FEP/C/FEP with SPC) or +260 °C continuous (PFA/C/PFA with SPC/NPC) with < 0.01 % moisture uptake
  • Dense >= 85 % plated Cu braid: Tinned, silver-plated or nickel-plated copper braid blocks VFD motor switching transients and 50/60 Hz common-mode hum on milliohm RTD bridges
  • Slim OD & 300 m factory reels: Ultra-compact 2.8 mm OD on 4x0.14 mm² probe tails up to 7.4 mm on 4x1.5 mm² field leads; MOQ 300 m continuous reels in 7 to 45 working days
Pt100 & Pt1000 Screened FEP & PFA RTD Sensor Cable (3-, 4- & 6-Wire) technical data
ParameterValue
Rated voltage300 V / 500 V AC instrumentation rating (2000 V AC 50 Hz dielectric test) with over 5 GOhm*km insulation resistance at 20 C
Circuit configurations and sizes3-wire, 4-wire (Kelvin) and 6-wire (duplex Pt100) in 0.14, 0.22, 0.25, 0.34, 0.50, 0.75, 1.00 and 1.50 mm2 (26 AWG to 16 AWG)
Fluoropolymer and plating optionsFEP/C/FEP (-90 C to +180 C TPC or +205 C SPC) and PFA/C/PFA (-90 C to +250 C SPC or +260 C NPC) with 85% plated copper braid
Sensor accuracy compatibilityLow inter-core resistance unbalance and zero moisture leakage preserve IEC 60751 Class AA, Class A and 1/10 DIN Pt100 / Pt1000 accuracy
Standard put-up and lead time300 m minimum order continuous factory reels per size and grade with a 7 to 45 working days manufacturing lead time
Core identificationIEC 60751 standard RTD polarity colours (red and white combinations; numbered or custom colour sets available for dual 4-wire 8-core probes)
Cleanroom and chemical rating100% silicone-free (PWIS/LABS-free), non-outgassing, FDA/USP-compatible fluoropolymer surface resistant to CIP/SIP chemicals and IEC 60332-1-2 flame

FEP/C/FEP (-90 °C to +205 °C) Screened Pt100 / Pt1000 RTD Cable Schedule

Screened FEP/C/FEP Pt100 / Pt1000 RTD Sensor Connection Cable (-90 °C to +205 °C) — Dimensional & Loop Resistance Schedule
Model & ConfigurationIEC 60751 Core ColoursStranding & PlatingMax. Cond. R at 20 °C (Ω/km)Nom. OD (mm)Copper Figure (kg/km)Approx. Weight (kg/km)Standard Reel MOQNext step
FEP/C/FEP 3x0.14 mm²2 Red + 1 White7x0.16 SPC138.02.68.514300 m reelRequest this size
FEP/C/FEP 4x0.14 mm²2 Red + 2 White7x0.16 SPC138.02.810.217300 m reelRequest this size
FEP/C/FEP 3x0.22 mm²2 Red + 1 White7x0.20 SPC85.02.912.420300 m reelRequest this size
FEP/C/FEP 4x0.22 mm²2 Red + 2 White7x0.20 SPC85.03.115.224300 m reelRequest this size
FEP/C/FEP 6x0.22 mm²4 Red + 2 White7x0.20 SPC85.03.821.534300 m reelRequest this size
FEP/C/FEP 3x0.34 mm²2 Red + 1 White7x0.25 SPC56.03.417.828300 m reelRequest this size
FEP/C/FEP 4x0.34 mm²2 Red + 2 White7x0.25 SPC56.03.721.633300 m reelRequest this size
FEP/C/FEP 3x0.50 mm²2 Red + 1 White19x0.18 SPC38.23.924.538300 m reelRequest this size
FEP/C/FEP 4x0.50 mm²2 Red + 2 White19x0.18 SPC38.24.330.045300 m reelRequest this size
FEP/C/FEP 4x0.75 mm²2 Red + 2 White19x0.23 SPC25.44.942.062300 m reelRequest this size
FEP/C/FEP 4x1.00 mm²2 Red + 2 White19x0.26 SPC19.15.454.078300 m reelRequest this size
FEP/C/FEP 4x1.50 mm²2 Red + 2 White19x0.32 SPC12.96.579.0112300 m reelRequest this size

PFA/C/PFA (-90 °C to +260 °C) Extreme-Temperature Pt100 RTD Cable Schedule

Screened PFA/C/PFA Extreme-Temperature Pt100 RTD Sensor Cable (-90 °C to +260 °C) — Dimensional Schedule
Model & ConfigurationIEC 60751 Core ColoursConductor & Braid PlatingContinuous Temp.Nom. OD (mm)Approx. Weight (kg/km)Standard Reel MOQNext step
Pt100 PFA/C/PFA 3x0.22 mm²2 Red + 1 White19x0.12 SPC or NPC-90 to +260 °C2.921300 m reelRequest this size
Pt100 PFA/C/PFA 4x0.22 mm²2 Red + 2 White19x0.12 SPC or NPC-90 to +260 °C3.125300 m reelRequest this size
Pt100 PFA/C/PFA 6x0.22 mm²4 Red + 2 White19x0.12 SPC or NPC-90 to +260 °C3.835300 m reelRequest this size
Pt100 PFA/C/PFA 3x0.50 mm²2 Red + 1 White19x0.18 SPC or NPC-90 to +260 °C3.939300 m reelRequest this size
Pt100 PFA/C/PFA 4x0.50 mm²2 Red + 2 White19x0.18 SPC or NPC-90 to +260 °C4.346300 m reelRequest this size
Pt100 PFA/C/PFA 4x0.75 mm²2 Red + 2 White19x0.23 SPC or NPC-90 to +260 °C4.964300 m reelRequest this size
Pt100 PFA/C/PFA 4x1.00 mm²2 Red + 2 White19x0.26 SPC or NPC-90 to +260 °C5.480300 m reelRequest this size

ASCII part references for BOM entry: TF-RTD-FEPC-3X022, TF-RTD-FEPC-4X022, TF-RTD-FEPC-6X022, TF-RTD-FEPC-4X050, TF-RTD-FEPC-4X100, TF-RTD-PFAC-4X022, TF-RTD-PFAC-4X050.

Note on conductor material: RTD connection cables ALWAYS use low-resistance plated copper conductors (never thermocouple alloys!), whereas thermocouple cables use K/J/T/E/N thermoelectric alloys.

Standards

Pt100 IEC 60751, DIN VDE 0207-6 & Flame Standards

Pt100 & Pt1000 Screened FEP & PFA RTD Sensor Cable (3-, 4- & 6-Wire) standards and test references
Standard or specificationApplication
IEC 60751 (DIN EN 60751)Industrial platinum resistance thermometers (Pt100, Pt500, Pt1000) accuracy classes (AA, A, B) and 2/3/4-wire red/white lead colour coding.
DIN VDE 0207-6 / ASTM B298 / ASTM B355Fluoropolymer insulation compounds and silver- or nickel-plated copper conductor stranding and DC resistance limits.
IEC 60332-1-2Single-cable vertical flame self-extinguishing test.

Certifications and document support

TEBAOFLEX ISO 9001 quality management and DC conductor resistance balance testing

IEC 60751 — Industrial platinum resistance thermometers and platinum temperature sensors

IEC 60332-1-2 — Vertical flame propagation test for insulated wires and cables

Compare by duty

Related cable options

Compare these cable options by service combination, movement, load, voltage and transmission requirement.

Engineering-led quotation

Request a cable recommendation and quotation

Send the required services, cable length, mechanical duty, environment, termination and quantity. Our team will review the application before quoting.

  • Include the product name or SKU plus the required cable length and quantity.
  • Attach a specification, drawing or data sheet when available.

    MOQ 50 m · Engineering reply within 24 hours

    By submitting this form, you agree that TEBAOFLEX may use the contact information and files provided to review and respond to your enquiry. Files are used for quotation and engineering communication related to your request. See our Privacy Policy for current data handling and retention information.

    Email or WhatsApp is enough. Please include the key cable or application details for review.